Medical device recalls Moderate risk

Fresenius Medical Care Renal Therapies Group, LLC recalls 2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy.

Recall date
October 5, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0829-2018
FDA classification
Class II
Brand / firm
Fresenius Medical Care Renal Therapies Group, LLC
Sold / distributed
US Nationwide Distribution

Why it was recalled

While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy.

Get recall alerts

Free email alert whenever Fresenius Medical Care Renal Therapies Group, LLC has a new recall — straight from official government data. Unsubscribe anytime.