Fresenius Medical Care Renal Therapies Group, LLC recalls 2008T HEMODIALYSIS SYS, WITH CDX, Part Number 190713, Applicable only with 2008T BlueStar Premium Upgrade Kit, Part Num…
- Recall date
- July 13, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-3194-2018
- FDA classification
- Class II
- Brand / firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Sold / distributed
- US Nationwide in the states of : CA, CT, MI, NC, and NY.
Why it was recalled
In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is powered up and a mandatory rinse should be required. However, if a power failure or interruption occurs during the cleaning/disinfection program with the auto-start feature enabled, the machine enters the dialysis program even though the dialysis program button is disabled and the mandatory rinse has not been performed. As a result, disinfecting chemical may not be completely removed from the machine before starting a treatment and disinfectant could exist in the dialysate delivered to the patient.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
2008T HEMODIALYSIS SYS, WITH CDX, Part Number 190713, Applicable only with 2008T BlueStar Premium Upgrade Kit, Part Number 191138 Product Usage: Indicated for acute and chronic dialysis therapy.