Medical device recalls Moderate risk

Fresenius Medical Care Renal Therapies Group, LLC recalls Fresenius 2008T Series Hemodialysis System

Recall date
December 21, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1026-2017
FDA classification
Class II
Brand / firm
Fresenius Medical Care Renal Therapies Group, LLC
Sold / distributed
Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam

Why it was recalled

When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Fresenius 2008T Series Hemodialysis System

Get recall alerts

Free email alert whenever Fresenius Medical Care Renal Therapies Group, LLC has a new recall — straight from official government data. Unsubscribe anytime.