Medical device recalls Moderate risk

Fresenius Modified Mixer Motor recalled over fire hazard

Recall date
October 23, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
Fresenius Medical Care Renal Therapies Group, LLC recalls Fresenius Modified Mixer Motor (for 132 Gallon Dissolution Units), P/N 160075. For Use in the Preparation of acid conce…
Recall number
Z-0296-2016
FDA classification
Class II
Brand / firm
Fresenius Medical Care Renal Therapies Group, LLC
Sold / distributed
Nationwide Distribution.

Why it was recalled

Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Fresenius Modified Mixer Motor (for 132 Gallon Dissolution Units), P/N 160075. For Use in the Preparation of acid concentrate for hemodialysis.

Get recall alerts

Free email alert whenever Fresenius Medical Care Renal Therapies Group, LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Fresenius Medical Care Renal Therapies Group, LLC