Fresenius Modified Mixer Motor recalled over fire hazard
- Recall date
- October 23, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Fresenius Medical Care Renal Therapies Group, LLC recalls Fresenius Modified Mixer Motor (for 132 Gallon Dissolution Units), P/N 160075. For Use in the Preparation of acid conce…
- Recall number
- Z-0296-2016
- FDA classification
- Class II
- Brand / firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Sold / distributed
- Nationwide Distribution.
Why it was recalled
Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Fresenius Modified Mixer Motor (for 132 Gallon Dissolution Units), P/N 160075. For Use in the Preparation of acid concentrate for hemodialysis.
Get recall alerts
Free email alert whenever Fresenius Medical Care Renal Therapies Group, LLC has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Fresenius Medical Care Renal Therapies Group, LLC