Medical device recalls Moderate risk

Fujifilm Medical Systems U.S.A., Inc. recalls FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name) The FDR Vision…

Recall date
June 15, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0232-2018
FDA classification
Class II
Brand / firm
Fujifilm Medical Systems U.S.A., Inc.
Sold / distributed
US Distribution.

Why it was recalled

A potential issue in the X-ray tube ceiling unit CH-200 which constitutes the Digital Radiography System FDR Visionary Suite. The X-ray tube assembly in these devices is mounted to a tube mounting flange, which is part of the tube holding shaft. where cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name) The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations.