Fujifilm Medical Systems U.S.A., Inc. recalls FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name) The FDR Vision…
- Recall date
- June 15, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0232-2018
- FDA classification
- Class II
- Brand / firm
- Fujifilm Medical Systems U.S.A., Inc.
- Sold / distributed
- US Distribution.
Why it was recalled
A potential issue in the X-ray tube ceiling unit CH-200 which constitutes the Digital Radiography System FDR Visionary Suite. The X-ray tube assembly in these devices is mounted to a tube mounting flange, which is part of the tube holding shaft. where cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name) The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations.
Get recall alerts
Free email alert whenever Fujifilm Medical Systems U.S.A., Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Fujifilm Medical Systems U.S.A., Inc.