Medical device recalls Moderate risk

Fujifilm Medical Systems U.S.A., Inc. recalls Fujifilm Medical Endoscope for the duodenum and upper G.I. tract. ED-530XT, ED-250XT5, ED-250XL5, ED-450XT5, and ED-450…

Recall date
June 30, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2778-2015
FDA classification
Class II
Brand / firm
Fujifilm Medical Systems U.S.A., Inc.
Sold / distributed
Distributed in the US and Canada.

Why it was recalled

The suction button on the ED-530XT endoscope may become lodged (stuck) in the active position following the use of x-ray contrast agents during ERCP procedures.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Fujifilm Medical Endoscope for the duodenum and upper G.I. tract. ED-530XT, ED-250XT5, ED-250XL5, ED-450XT5, and ED-450XL5 Endoscope Operation Manuals.