Fujifilm Medical Systems U.S.A., Inc. recalls Synapse Cardiovascular (Synapse CV). Software Versions: 6.0 to 6.2.1 The Synapse Cardiovascular image is a management a…
- Recall date
- April 30, 2021
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1775-2021
- FDA classification
- Class II
- Brand / firm
- Fujifilm Medical Systems U.S.A., Inc.
- Sold / distributed
- Worldwide distribution - US Nationwide distribution and the countries of Canada, Bermuda, Israel, Hadassah Medical Center.
Why it was recalled
Software versions 6.0.4 to 6.2.1 using Advanced Reporting -possibility that a previously assigned internal patient database ID can be reused for a new patient. It was discovered that this value does not increase when a patient merge activity is executed in a specific sequence
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Synapse Cardiovascular (Synapse CV). Software Versions: 6.0 to 6.2.1 The Synapse Cardiovascular image is a management and reporting system