Medical device recalls Moderate risk

Ge Healthcare It recalls Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. Qualified sys…

Recall date
November 16, 2012
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2739-2016
FDA classification
Class II
Brand / firm
Ge Healthcare It
Sold / distributed
Nationwide Distribution

Why it was recalled

There is a potential for data loss associated with empty path names resulting from HL-7 updates when the cache disk is almost full. In an unlikely event that the cache disk is almost full (<100MB) and HL-7 update events are received, the update action fails and an empty path name is written to the EA database. As a result the instance is not deleted from the storage device but no longer referenced. When trying to retrieve the instance an error is returned.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. Qualified system administrators install, monitor and maintain the system. DICOM devices (e.g. modalities, workstations) communicate with the archive using the DICOM protocol (published by ACR-NEMA). The clinical use is limited to the interaction other systems have with the Centricity Enterprise Archive. Centricity Enterprise Archive is not intended for use in patient vicinity. The intended users for Centricity Enterprise Users service personnel and System administrators, and non clinical users. As of V4.0 XDS enabled systems can communicate with the archive using the XDS and XDS-I profiles (published by IHE).

Get recall alerts

Free email alert whenever Ge Healthcare It has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Ge Healthcare It