Medical device recalls Moderate risk

GE Healthcare It recalls GE Centricity PACS Workstation, versions 3.1.1.x through 3.2.1.x Picture Archiving and Communication System Used as a p…

Recall date
November 15, 2012
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2579-2016
FDA classification
Class II
Brand / firm
GE Healthcare It
Sold / distributed
Worldwide distribution including US nationwide, Puerto Rico, Australia, Austria, Belgium, Bermuda, CANADA, Denmark, Egypt, ENGLAND, France, Germany, India, Ireland, Hong Kong, Korea, Kuwait, Italy, Malta, Netherlands, Portugal, Qatar, Saudi Arabia, SCOTLAND, South Africa, South Korea, Spain, Sweden…

Why it was recalled

While merging exams in a test instance, merges carry forward in the production system, when the middle tier is configured to the wrong IP address of the Centricity Exam Manager. When a current and historical exam are opened at the same time, the system asynchronously tries to access the operation specifying how each image should be grouped, causing random image-series grouping errors.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

GE Centricity PACS Workstation, versions 3.1.1.x through 3.2.1.x Picture Archiving and Communication System Used as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. Also used as a clinical review workstation throughout the healthcare facility. The workstation interface provides the user with a means to display, manipulate, archive, print and export images when connected with the Centricity PACS infrastructure.

Get recall alerts

Free email alert whenever GE Healthcare It has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: GE Healthcare It