Medical device recalls Moderate risk

GE Healthcare, LLC recalls Aisys Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.…

Recall date
September 27, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0411-2019
FDA classification
Class II
Brand / firm
GE Healthcare, LLC
Sold / distributed
Worldwide distribution - US nationwide in the states of CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, ,MA, MD, MI, MN, MO, MT, NC, NE, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. Countries of Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, China, Colombia, Croatia, Cz…

Why it was recalled

Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Aisys Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH FOR USA, AISYS 10.X TO 11.X UPGRADE KIT FRENCH, AISYS 10.X TO 11.X UPGRADE KIT GERMAN, AISYS 10.X TO 11.X UPGRADE KIT JAPANESE.

Get recall alerts

Free email alert whenever GE Healthcare, LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: GE Healthcare, LLC