Medical device recalls Moderate risk

Infant Warmer System recalled over fire hazard

Recall date
August 4, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
GE Healthcare, LLC recalls Infant Warmer System (IWS)
Recall number
Z-1015-2017
FDA classification
Class II
Brand / firm
GE Healthcare, LLC
Sold / distributed
This product has been distributed domestically and internationally .

Why it was recalled

GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Infant Warmer System (IWS)

Get recall alerts

Free email alert whenever GE Healthcare, LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: GE Healthcare, LLC