Infant Warmer System recalled over fire hazard
- Recall date
- May 5, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- GE Healthcare, LLC recalls Infant Warmer System (IWS)
- Recall number
- Z-2323-2017
- FDA classification
- Class II
- Brand / firm
- GE Healthcare, LLC
- Sold / distributed
- Worldwide Distribution - USA (nationwide) Distribution.
Why it was recalled
GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Infant Warmer System (IWS)