GE Healthcare, LLC recalls NM/CT 870 CZT, Model/Catalogue/Code # H3906CW Product Usage: The GE Discovery NM/CT 670 CZT system is intended for gene…
- Recall date
- March 20, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1264-2019
- FDA classification
- Class II
- Brand / firm
- GE Healthcare, LLC
- Sold / distributed
- US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK
Why it was recalled
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
NM/CT 870 CZT, Model/Catalogue/Code # H3906CW Product Usage: The GE Discovery NM/CT 670 CZT system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using a variety of scanning modes supported by various acquisition types, and a Computed Tomography system which is intended for enabling attenuation correction and anatomical localization of SPECT images and for standalone head, whole body, cardiac and vascular X-ray Computed Tomography applications.