GE Healthcare, LLC recalls Revolution CT systems with the SmartStep Option The system is intended for head, whole body, cardiac and vascular X-ray…
- Recall date
- July 12, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0776-2019
- FDA classification
- Class II
- Brand / firm
- GE Healthcare, LLC
- Sold / distributed
- Worldwide Distribution - US (Nationwide Distribution) and countries of: Australia, Belgium, Canada, Estonia, Finland, France, Germany, Hong Kong, Ireland, Japan, Korea Republic of, Malaysia, Morocco, New Zealand, Norway, Saudi Arabia, Singapore, Slovakia, Sweden, Switzerland, Thailand, Turkey, Unit…
Why it was recalled
On the Revolution CT systems equipped with the SmartStep Option, the Z location displayed on image annotation and on the gantry display may not be the actual location of the table and result in the system scanning at a location that is not what was confirmed by the operator
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Revolution CT systems with the SmartStep Option The system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications.