Medical device recalls Moderate risk

GE Healthcare, LLC recalls Uterine Electromyographic Monitor -Monica IF24 Interface System Product Usage: The Monica IF24 CTG Interface Device is…

Recall date
December 15, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1293-2018
FDA classification
Class II
Brand / firm
GE Healthcare, LLC
Sold / distributed
Worldwide - US Nationwide Distribution.

Why it was recalled

if the cables of the Monica IF24 System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Uterine Electromyographic Monitor -Monica IF24 Interface System Product Usage: The Monica IF24 CTG Interface Device is an accessory to the Monica AN24 which provides a means of interfacing the wireless output of the Monica AN24 to the transducer inputs of a CTG Fetal monitor. The Monica IF24 CTG Interface Device enables signals collected by the Monica AN24 to be printed and displayed on a CTG Fetal Monitor and sent on to a central network, if connected. The device is intended for use by healthcare professionals in a clinical setting.

Get recall alerts

Free email alert whenever GE Healthcare, LLC has a new recall — straight from official government data. Unsubscribe anytime.