GE Healthcare, LLC recalls Uterine Electromyographic Monitor - Monica Novii Wireless Patch System Product Usage: The Novii Interface is an accesso…
- Recall date
- December 15, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1292-2018
- FDA classification
- Class II
- Brand / firm
- GE Healthcare, LLC
- Sold / distributed
- Worldwide - US Nationwide Distribution.
Why it was recalled
if the cables of the Monica Novii System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Uterine Electromyographic Monitor - Monica Novii Wireless Patch System Product Usage: The Novii Interface is an accessory to the Novii Pod which provides a means of interfacing the wireless output of the Novii Pod to the transducer inputs of a CTG Fetal monitor via three cables. The Novii Interface enables signals collected by the Novii Pod to be printed and displayed on a CTG Fetal Monitor and sent on to a central network, if connected.
Get recall alerts
Free email alert whenever GE Healthcare, LLC has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: GE Healthcare, LLC