GE Healthcare, LLC recalls Vivid S60/S70/S60N/S70N, H45041SU, H45041SW, H45581MS, H45581PD.
- Recall date
- March 10, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1780-2017
- FDA classification
- Class II
- Brand / firm
- GE Healthcare, LLC
- Sold / distributed
- United Kingdom, United Arab Emirates, Turkey, Thailand, Switzerland, Sweden, Spain, South Africa, Slovenia, Saudi Arabia, Russia, Romania, Portugal, Poland, Philippines, Norway, New Zealand, New Caledonia, Netherlands, Myanmar, Kuwait, Korea (Republic of), Japan, Italy, ISRAEL, Ireland, Iran, India…
Why it was recalled
GE Healthcare has become aware of an issue where a patient other than the intended is incorrectly selected by the operator in situations where DICOM Worklist search response time is slow. This issue is limited to certain Vivid Ultrasound systems and can result in the incorrect patient information showing on the screen during the exam. If not detected by the operator, images from the actual (intended) patient will be stored under the incorrect patient after the exam. This issue could lead to misdiagnosis. There have been no injuries reported because of this issue.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Vivid S60/S70/S60N/S70N, H45041SU, H45041SW, H45581MS, H45581PD.