Medical device recalls Moderate risk

GE Medical Systems, LLC recalls GE Healthcare, Aespire 7900, 1009-9012-000 & Aespire View, 1009-9212-000. The family of GE Datex-Ohmeda Aespire anesthe…

Recall date
December 15, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0677-2016
FDA classification
Class II
Brand / firm
GE Medical Systems, LLC
Sold / distributed
Worldwide Distribution - US (nationwide including PR except NH, VT, & WY. Internationally to ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BOLIVIA, BOSNIA, BRAZIL, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ESTONIA, FINLAND…

Why it was recalled

The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

GE Healthcare, Aespire 7900, 1009-9012-000 & Aespire View, 1009-9212-000. The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The devices are intended for volume or pressure control ventilation.