GE OEC Medical Systems, Inc recalls OEC 9900 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 9900 Elite mobile fluoroscopy system is desig…
- Recall date
- November 21, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0705-2025
- FDA classification
- Class II
- Brand / firm
- GE OEC Medical Systems, Inc
- Sold / distributed
- U.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden,…
Why it was recalled
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
OEC 9900 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 9900 Elite mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the patient during diagnostic surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.