Medical device recalls Moderate risk

Gentian AS recalls GENTIAN Cystatin C Calibrator and Calibrator Kit, for the following models: 1. Gentian C E, Cystatin C Calibrator, 1 x…

Recall date
September 2, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0099-2021
FDA classification
Class II
Brand / firm
Gentian AS
Sold / distributed
US: MN, VA, KS,

Why it was recalled

The firm has identified lots of calibrators that have shown to measure incorrect high values over time compared to the assigned values when released for distribution. This could result in higher Cystatin C values measured in patients.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

GENTIAN Cystatin C Calibrator and Calibrator Kit, for the following models: 1. Gentian C E, Cystatin C Calibrator, 1 x 1 mL, REF 1012, UDI: (01)07090038460092 2. Gentian C E Calibrator Kit, 6 x 1 mL, REF 1051, UDI: (01)07090038460108 3. Gentian C E, BCI Ref: A52763, Cystatin C calibrator Kit, 6 x 1 mL, REF 1051, (01)17090038460037 4. Gentian C E, Cystatin C Calibrator, 1 x 1 mL, REF 8012 (Not distributed in US)