Getinge 88-Series Washer- Disinfector recalled over fire hazard
- Recall date
- August 29, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Getinge Disinfection Ab recalls Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
- Recall number
- Z-0016-2026
- FDA classification
- Class II
- Brand / firm
- Getinge Disinfection Ab
- Sold / distributed
- Worldwide - US Nationwide distribution in the states of Colorado, Massachusetts, New Mexico, New Hampshire and the countries of Algeria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Austria, Hong Kong, Hungary, India, Iraq,…
Why it was recalled
Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Getinge 88-Series Washer- Disinfector. Model Number: 88-5.