Getinge Disinfection Ab recalls Washer Disinfector Aquadis 56; Model Number: 56M
- Recall date
- April 2, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1721-2025
- FDA classification
- Class II
- Brand / firm
- Getinge Disinfection Ab
- Sold / distributed
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Brazil, Canada, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran,…
Why it was recalled
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Washer Disinfector Aquadis 56; Model Number: 56M
Get recall alerts
Free email alert whenever Getinge Disinfection Ab has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Getinge Disinfection Ab