Medical device recalls Moderate risk

Getinge Group Logistics America, LLC recalls Operating Table System

Recall date
February 11, 2021
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1296-2021
FDA classification
Class II
Brand / firm
Getinge Group Logistics America, LLC
Sold / distributed
Domestic: AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV. Foreign: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colom…

Why it was recalled

There is a potential for hydraulic leakage at the inclination function to occur resulting in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning which may lead to the patient having to be transferred to another OR table causing a delay in the procedure.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Operating Table System