Medical device recalls Moderate risk

GRAMERCY EXTREMITY ORTHOPEDICS recalls GEO Instrument Kit for EPS Plate, REF: 70120001, part of the GEO FirstFUSE 1st MTP Joint Arthrodesis Plating System

Recall date
April 4, 2023
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1395-2023
FDA classification
Class II
Brand / firm
GRAMERCY EXTREMITY ORTHOPEDICS
Sold / distributed
US Nationwide distribution in the states of CT, OR, NJ, AZ, CT, OH, AZ.

Why it was recalled

Instrument Kits for EPS plates may contain drill guides with incorrect drill depth range of 30mm - 60mm; not the correct drill depth range of 10mm - 40mm.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

GEO Instrument Kit for EPS Plate, REF: 70120001, part of the GEO FirstFUSE 1st MTP Joint Arthrodesis Plating System

Get recall alerts

Free email alert whenever GRAMERCY EXTREMITY ORTHOPEDICS has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: GRAMERCY EXTREMITY ORTHOPEDICS