GRAMERCY EXTREMITY ORTHOPEDICS recalls GEO Instrument Kit for EPS Plate, REF: 70120001, part of the GEO FirstFUSE 1st MTP Joint Arthrodesis Plating System
- Recall date
- April 4, 2023
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1395-2023
- FDA classification
- Class II
- Brand / firm
- GRAMERCY EXTREMITY ORTHOPEDICS
- Sold / distributed
- US Nationwide distribution in the states of CT, OR, NJ, AZ, CT, OH, AZ.
Why it was recalled
Instrument Kits for EPS plates may contain drill guides with incorrect drill depth range of 30mm - 60mm; not the correct drill depth range of 10mm - 40mm.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
GEO Instrument Kit for EPS Plate, REF: 70120001, part of the GEO FirstFUSE 1st MTP Joint Arthrodesis Plating System