Medical device recalls Moderate risk

Greatbatch Medical recalls Torque Limiting Adapter: Torque Limiting Attachment, Non Sterile, Distributed by Zimmer. PRECIMED, Torque Limiting Adap…

Recall date
April 26, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1884-2016
FDA classification
Class II
Brand / firm
Greatbatch Medical
Sold / distributed
US: IN, MI, TX. OUS: GERMANY, FRANCE, ITALY, SWITZERLAND.

Why it was recalled

Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Torque Limiting Adapter: Torque Limiting Attachment, Non Sterile, Distributed by Zimmer. PRECIMED, Torque Limiting Adaptor, Non Sterile.. Customer Number/Model: 00-2360-080-00 T10232, T116245 T116245, TLA025-25 TLA025-25, 00-2360-080-05 T10233, MAC21003 T116430, TLA025-30 TLA025-30, MAC21001 T112972, MAC21001 T13621, TLA025-20 TLA025-20.

Get recall alerts

Free email alert whenever Greatbatch Medical has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Greatbatch Medical