Medical device recalls Moderate risk

Halyard Health, Inc recalls CORTRAK* 2 Nasogastric/Nasointestinal (NG/NI) Feeding Tubes with Electromagnetic Transmitting Stylet and CORFLO* Entera…

Recall date
December 13, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1277-2018
FDA classification
Class II
Brand / firm
Halyard Health, Inc
Sold / distributed
Worldwide Distribution: US (nationwide) to states of: AR, CA, CO, GA, IA, LA, MI, MN, NM, TX; and country of Germany. Distributed internationally to Germany.

Why it was recalled

The transmitting stylet within CORTRAK* 2 nasogastric/Nasointestinal (NG/NI) Feeding Tubes with ANTI-IV* connector (Product Code 20-9551Trak2A) may be visible within the window of the exit port of the feeding tube. Based on specifications, the transmitting stylet in the NG tube should not be visible within the port window.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

CORTRAK* 2 Nasogastric/Nasointestinal (NG/NI) Feeding Tubes with Electromagnetic Transmitting Stylet and CORFLO* Enteral Feeding tube, ANTI-IV* Connector, 10 Fr, 140 cm The device is used in patients that require intermittent or continuous tube feedings via the nasogastric or nasoenteric route.