Halyard Health, Inc recalls CORTRAK* 2 Nasogastric/Nasointestinal (NG/NI) Feeding Tubes with Electromagnetic Transmitting Stylet and CORFLO* Entera…
- Recall date
- December 13, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1277-2018
- FDA classification
- Class II
- Brand / firm
- Halyard Health, Inc
- Sold / distributed
- Worldwide Distribution: US (nationwide) to states of: AR, CA, CO, GA, IA, LA, MI, MN, NM, TX; and country of Germany. Distributed internationally to Germany.
Why it was recalled
The transmitting stylet within CORTRAK* 2 nasogastric/Nasointestinal (NG/NI) Feeding Tubes with ANTI-IV* connector (Product Code 20-9551Trak2A) may be visible within the window of the exit port of the feeding tube. Based on specifications, the transmitting stylet in the NG tube should not be visible within the port window.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
CORTRAK* 2 Nasogastric/Nasointestinal (NG/NI) Feeding Tubes with Electromagnetic Transmitting Stylet and CORFLO* Enteral Feeding tube, ANTI-IV* Connector, 10 Fr, 140 cm The device is used in patients that require intermittent or continuous tube feedings via the nasogastric or nasoenteric route.
Get recall alerts
Free email alert whenever Halyard Health, Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Halyard Health, Inc