Hamilton Medical, Inc. recalls Hamilton Medical Infant Flow Sensor, single use, (1.6m), Part number 260177. The infant flow sensor is an accessory for…
- Recall date
- June 1, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1963-2015
- FDA classification
- Class II
- Brand / firm
- Hamilton Medical, Inc.
- Sold / distributed
- US Nationwide Distribution.
Why it was recalled
An issue has been discovered with the Hamilton Medical Infant Flow Sensor, single use. When a ventilation mode with adaptive volume control (APVcmv/CMV+) is used, the ventilator may adjust the ventilation on the inaccurate flow measurement and decrease the amount of ventilation delivered.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Hamilton Medical Infant Flow Sensor, single use, (1.6m), Part number 260177. The infant flow sensor is an accessory for the Hamilton Medical Ventilators. Product Usage: The infant flow sensor is a device to measure patient air flow and pressures within the Hamilton Medical Ventilators.