Medical device recalls Moderate risk

Handicare Usa Inc recalls PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered

Recall date
September 8, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0013-2018
FDA classification
Class II
Brand / firm
Handicare Usa Inc
Sold / distributed
Us and Canada

Why it was recalled

During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered