Handicare Usa Inc recalls PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered
- Recall date
- September 8, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0013-2018
- FDA classification
- Class II
- Brand / firm
- Handicare Usa Inc
- Sold / distributed
- Us and Canada
Why it was recalled
During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered
Get recall alerts
Free email alert whenever Handicare Usa Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Handicare Usa Inc