Medical device recalls Moderate risk

Healgen Scientific Llc recalls Healgen Series Reagent strips for Urinalysis are composed of several color pads aligned on a test strip. Each pad is em…

Recall date
August 22, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-3303-2024
FDA classification
Class II
Brand / firm
Healgen Scientific Llc
Sold / distributed
U.S. Nationwide distribution in the states of CA, GA and TX.

Why it was recalled

The reason for this recall is Healgen Scientific Ltd. has determined that their Urinary Reagent Strip products have shown reduced sensitivity, or potential false negative results for the detection of urinary protein by way of uneven color fading on the total protein test pad. This could lead to a delay in the identification of potential kidney diseases, their evaluation, and/or other conditions associated with elevated protein levels in urine and subsequently a delay in patient treatment.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Healgen Series Reagent strips for Urinalysis are composed of several color pads aligned on a test strip. Each pad is employed for testing one assay item by visually or instrumentally reading the color change of the pad and comparing with the corresponding blocks on a color chart. The Healgen Series Reagent Strips for Urinalysis provides tests for Glucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Ascorbic Acid and Leukocytes in Urine.

Get recall alerts

Free email alert whenever Healgen Scientific Llc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Healgen Scientific Llc