Healgen Scientific Llc recalls Healgen Series Reagent strips for Urinalysis are composed of several color pads aligned on a test strip. Each pad is em…
- Recall date
- August 22, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-3303-2024
- FDA classification
- Class II
- Brand / firm
- Healgen Scientific Llc
- Sold / distributed
- U.S. Nationwide distribution in the states of CA, GA and TX.
Why it was recalled
The reason for this recall is Healgen Scientific Ltd. has determined that their Urinary Reagent Strip products have shown reduced sensitivity, or potential false negative results for the detection of urinary protein by way of uneven color fading on the total protein test pad. This could lead to a delay in the identification of potential kidney diseases, their evaluation, and/or other conditions associated with elevated protein levels in urine and subsequently a delay in patient treatment.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Healgen Series Reagent strips for Urinalysis are composed of several color pads aligned on a test strip. Each pad is employed for testing one assay item by visually or instrumentally reading the color change of the pad and comparing with the corresponding blocks on a color chart. The Healgen Series Reagent Strips for Urinalysis provides tests for Glucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Ascorbic Acid and Leukocytes in Urine.
Get recall alerts
Free email alert whenever Healgen Scientific Llc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Healgen Scientific Llc