Hill-Rom, Inc. recalls FlexoStretch Model Number: 3156057 - The combination of this products allows the user to easily adjust/control the cent…
- Recall date
- August 11, 2021
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2598-2021
- FDA classification
- Class II
- Brand / firm
- Hill-Rom, Inc.
- Sold / distributed
- Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, DE, IL, ME, MI, MO, MS, NJ, OH, PA, VA, WA and the countries of Argentina, Australia, Austria, Bolivia, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Gr…
Why it was recalled
Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide backwards and potentially may fall out of the Oct lift sheet back onto the surface it is being lifted from.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
FlexoStretch Model Number: 3156057 - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient.
Get recall alerts
Free email alert whenever Hill-Rom, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Hill-Rom, Inc.