Hill-Rom, Inc. recalls LikoStretch Mod 600 IC Model Number: 3156065 LikoStretch Mod 600 IC, Wide Model Number: 3156065B (obsolete since 12JUN2…
- Recall date
- August 11, 2021
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2601-2021
- FDA classification
- Class II
- Brand / firm
- Hill-Rom, Inc.
- Sold / distributed
- Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, DE, IL, ME, MI, MO, MS, NJ, OH, PA, VA, WA and the countries of Argentina, Australia, Austria, Bolivia, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Gr…
Why it was recalled
Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide backwards and potentially may fall out of the Oct lift sheet back onto the surface it is being lifted from.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
LikoStretch Mod 600 IC Model Number: 3156065 LikoStretch Mod 600 IC, Wide Model Number: 3156065B (obsolete since 12JUN2020) - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient.