Medical device recalls Moderate risk

Hospira Inc. recalls LifeCare PCA infusion system with Hospira MedNet software allows clinicians to administer, or patients to self-administ…

Recall date
March 8, 2013
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2141-2016
FDA classification
Class II
Brand / firm
Hospira Inc.
Sold / distributed
Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY) + District of Columbia + Puerto Rico + Virgin Islands; *** FOREIGN:

Why it was recalled

Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

LifeCare PCA infusion system with Hospira MedNet software allows clinicians to administer, or patients to self-administer analgesia, safely and effectively within clinician programmed limits and/or hospital-defined medication limits. The LifeCare PCA infusion system is used in a wide range of clinical settings that includes but is not limited to the following: medical, laboratory/delivery/post-partum, burn unit, surgical, operating room, oncology, critical care units, post-anesthesia care unit (PACU), pediatrics.