Medical device recalls Moderate risk

Hospira Inc. recalls The LifeCare PCA Plus II infusion pump system is designed for safe, effective, analgesic delivery to patients through t…

Recall date
March 8, 2013
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2139-2016
FDA classification
Class II
Brand / firm
Hospira Inc.
Sold / distributed
Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY) + District of Columbia + Puerto Rico + Virgin Islands; *** FOREIGN:

Why it was recalled

Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

The LifeCare PCA Plus II infusion pump system is designed for safe, effective, analgesic delivery to patients through the intravenous or epidural routes in a wide range of clinical settings, including ICU, CCU, SICU, trauma units, bone marrow transplant centers, burn units, oncology centers, and medical/surgical nursing units.