Imactis recalls Imactis Navigation Kit, REF 10100, Content Sterile Needle Holder Ref. A10010 CE 0459, Sensor Cover Ref. PC3688 CE 0482,…
- Recall date
- August 1, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0838-2020
- FDA classification
- Class II
- Brand / firm
- Imactis
- Sold / distributed
- US: WI & IL
Why it was recalled
The firm has become aware that the sensor cover, a component of the single-use navigation kit may contain microbial contamination. A source of microbial contamination has been identified in a raw product used to manufacture the tip of the cover.This could result in direct exposure of the intra-cardiovascular, intra-lymphatic or neurological system to a level of endotoxin sufficient to produce a pyrogenic response.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Imactis Navigation Kit, REF 10100, Content Sterile Needle Holder Ref. A10010 CE 0459, Sensor Cover Ref. PC3688 CE 0482, Adhesif Film Ref. 38.20.15 CE 0459, EN Navigation Kit. Compatible with linear instruments of diameter ranging from 0.5 mm to 3.0 mm (25G to 11G), Sterile, Rx Only,
Get recall alerts
Free email alert whenever Imactis has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Imactis