Implant Direct Sybron Manufacturing, LLC recalls 6534-08N InterActive¿ Implant Open-Tray Transfer Narrow Part Number: 6534-08N Product Usage: Intended Use of Implant Di…
- Recall date
- May 19, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2046-2016
- FDA classification
- Class II
- Brand / firm
- Implant Direct Sybron Manufacturing, LLC
- Sold / distributed
- Worldwide Distribution - U.S. Nationwide in the states of OR, NE, VA, FL, MO, WA, TX and the countries of CZ, GB, NL, MF, DK, IT, IE, DE
Why it was recalled
The incorrect sized product, InterActive Implant Open-Tray Transfer Wide, 6534-08W, was packaged instead of the InterActive Implant Open-tray Transfer Narrow, 6534-08N, causing a possible issue with the impression. This discrepancy may lead to the possibility of an oversized crown fabrication.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
6534-08N InterActive¿ Implant Open-Tray Transfer Narrow Part Number: 6534-08N Product Usage: Intended Use of Implant Direct Open Tray Transfers are to capture implant position in an elastomeric impression of the mandible or maxilla. Open Tray transfers stay within the Impression material prior to delivery to the laboratory technician.
Get recall alerts
Free email alert whenever Implant Direct Sybron Manufacturing, LLC has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Implant Direct Sybron Manufacturing, LLC