Medical device recalls High risk

Infusion Pump Repair recalls Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part

Recall date
March 26, 2021
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1544-2021
FDA classification
Class I
Brand / firm
Infusion Pump Repair
Sold / distributed
California

Why it was recalled

Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part