Infusion Pump Repair recalls Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part
- Recall date
- March 26, 2021
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1544-2021
- FDA classification
- Class I
- Brand / firm
- Infusion Pump Repair
- Sold / distributed
- California
Why it was recalled
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part
Get recall alerts
Free email alert whenever Infusion Pump Repair has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Infusion Pump Repair