Medical device recalls High risk

InfuTronix LLC recalls Nimbus Administration Set; Nimbus Flex; Nimbus PainPro (a.k.a. Halo Ambulatory Infusion System); Catalog Number IT1025.…

Recall date
February 21, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1285-2024
FDA classification
Class I
Brand / firm
InfuTronix LLC
Sold / distributed
Domestic US distribution nationwide. No international distribution.

Why it was recalled

InfuTronix is removing the Nimbus Infusion Pump Systems from the market due to multiple potential failure modes, including battery failure, upstream occlusion, system errors, drug product leakage, high or low flow rate, or damaged pump housing. The device will not be available or supported after June 20, 2024.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Nimbus Administration Set; Nimbus Flex; Nimbus PainPro (a.k.a. Halo Ambulatory Infusion System); Catalog Number IT1025. Ambulatory Infusion Pump and components

Get recall alerts

Free email alert whenever InfuTronix LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: InfuTronix LLC