Medical device recalls Moderate risk

INNOKAS MEDICAL OY recalls VC150 Vital Signs Monitor (VC150 monitor equipped with Masimo SpO2 technique only); Innokas Medical CARESCAPE VC150 Rx…

Recall date
December 3, 2014
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1022-2015
FDA classification
Class II
Brand / firm
INNOKAS MEDICAL OY
Sold / distributed
Distributed in the US and Europe.

Why it was recalled

If the SpO2 or RRa value is violating the respective alarm limit at the time of switching to monitoring mode, the monitor does not issue visual and audible alarms.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

VC150 Vital Signs Monitor (VC150 monitor equipped with Masimo SpO2 technique only); Innokas Medical CARESCAPE VC150 Rx Only Innokas Medical Oy. Catalog numbers 2067980-002, 2067980-006, 2067980-010, and 2067980-014.

Get recall alerts

Free email alert whenever INNOKAS MEDICAL OY has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: INNOKAS MEDICAL OY