Medical device recalls Moderate risk

Inpeco S.A. recalls FlexLab Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Mod…

Recall date
April 7, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2140-2020
FDA classification
Class II
Brand / firm
Inpeco S.A.
Sold / distributed
US Nationwide including in the states of IL, NY.

Why it was recalled

The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of these Interface Modules is put off-line after a carrier in their secondary lane is physically returned on the main track and then put back on-line when the carrier is used to transport another tube.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

FlexLab Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Module (P/N FLX-226-01, FLX-226-10), or StaRRsed Interface Module (P/N FLX-268-00)is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.