Insulet Corporation recalls Omnipod DASH Insulin Management System (mg/dL configuration), Catalog Number PT000010 M/D: INT1D001MG - Product Usage:…
- Recall date
- April 2, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2164-2020
- FDA classification
- Class II
- Brand / firm
- Insulet Corporation
- Sold / distributed
- The products were distributed to the following foreign countries: Italy, Netherlands, UK.
Why it was recalled
After the device has been in use for about 2 months, data processing in the PDM can be slowed such that the Bolus Calculator fails to accurately subtract the correct amout of IOB before suggesting a bolus amount.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Omnipod DASH Insulin Management System (mg/dL configuration), Catalog Number PT000010 M/D: INT1D001MG - Product Usage: is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.
Get recall alerts
Free email alert whenever Insulet Corporation has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Insulet Corporation