Medical device recalls Moderate risk

Integra LifeSciences Corp. recalls MAYFIELD SPINE TABLE ADAPTOR RADIOLUCENT, Part Number A2600R The MAYFIELD¿ Infinity XR2 Base Unit is intended to be use…

Recall date
June 29, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2595-2018
FDA classification
Class II
Brand / firm
Integra LifeSciences Corp.
Sold / distributed
Worldwide Distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV; and to countries of:: Australia, Canada, China, Dominican Rep…

Why it was recalled

There is a potential for the knob assembly in the swivel adapter to fracture/break during use.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

MAYFIELD SPINE TABLE ADAPTOR RADIOLUCENT, Part Number A2600R The MAYFIELD¿ Infinity XR2 Base Unit is intended to be used to support a patient during diagnostic examination and/or surgical procedures where a rigid support between surgical table and headrest, or skull clamp is necessary, positional freedom is required and where X-ray imaging modalities will be used.