Integra LifeSciences Corp. recalls Universal Compact Head Ring Adapter Plate (UCHRAP), a component of the Universal Compact Head Ring Adapter, UCHRA. The…
- Recall date
- August 21, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0025-2016
- FDA classification
- Class II
- Brand / firm
- Integra LifeSciences Corp.
- Sold / distributed
- Worldwide Distribution - US Distribution to the states of : NJ, TX and WY., and to the countries of : Algeria, Germany and India.
Why it was recalled
Integra identified that the UCHRAP component (Arc Adapter Plate) would not assemble properly to the UCHRAR component (ARC Adapter Ring) of the Compact Head Ring Adapter due to an error in the manufacturing drawing.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Universal Compact Head Ring Adapter Plate (UCHRAP), a component of the Universal Compact Head Ring Adapter, UCHRA. The UCHRA is a component of the Integra¿ CRW¿ Precision Arc Stereotactic System.