Da Vinci Surgical System recalled over fire hazard
- Recall date
- June 17, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Intuitive Surgical, Inc. recalls da Vinci Surgical System, IS 4000; Endoscope Controller, Model Number: 372601-16 (UDI: 00886874114193) is included in t…
- Recall number
- Z-0283-2021
- FDA classification
- Class II
- Brand / firm
- Intuitive Surgical, Inc.
- Sold / distributed
- US Nationwide distribution including in the state of Washington.
Why it was recalled
One da Vinci Xi Endoscope Controller in the field was improperly calibrated during servicing. The result of this improper calibration may result in the tip of the endoscope may reach higher temperature causing the potential for thermal tissue injury and/or overly bright images.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
da Vinci Surgical System, IS 4000; Endoscope Controller, Model Number: 372601-16 (UDI: 00886874114193) is included in the IS 4000 Vision Cart, Model Number: 381121-31 (UDI 00886874110898) - Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.
Get recall alerts
Free email alert whenever Intuitive Surgical, Inc. has a new recall — straight from official government data. Unsubscribe anytime.