Medical device recalls Moderate risk

Intuitive Surgical, Inc. recalls da Vinci Xi Surgical System Product Usage: The da Vinci Xi Surgical System (IS4000) is designed to assist a surgeon in…

Recall date
June 1, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2630-2017
FDA classification
Class II
Brand / firm
Intuitive Surgical, Inc.
Sold / distributed
Worldwide Distribution - US Nationwide in the states of TX, NY,FL, IL, OH, and countries of France, Ireland, Japan, and Taiwan

Why it was recalled

da Vinci Xi(R) Surgical System Patient Side Carts (PSC) were shipped to the field with potentially under-torqued brake screws in a section of the system arms.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

da Vinci Xi Surgical System Product Usage: The da Vinci Xi Surgical System (IS4000) is designed to assist a surgeon in the accurate control of surgical instruments such as endoscopes, scissors, forceps, and retractors during endoscopic surgery. The system has four universal manipulators (arms) which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures in the pelvic, abdominal, and thoracic cavities.