Intuitive Surgical, Inc. recalls Ion Catheter Reprocessing Cover, IF1000 (Part Number:490114-02) UDI:00886874116845.
- Recall date
- June 10, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2949-2020
- FDA classification
- Class II
- Brand / firm
- Intuitive Surgical, Inc.
- Sold / distributed
- U.S.: California, Colorado, Florida, Illinois, Massachusetts, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Texas, and Virginia. No foreign consignees.
Why it was recalled
One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Ion Catheter Reprocessing Cover, IF1000 (Part Number:490114-02) UDI:00886874116845.
Get recall alerts
Free email alert whenever Intuitive Surgical, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Intuitive Surgical, Inc.