Ion Beam Applications S.A. recalls Proteus 235
- Recall date
- February 9, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2592-2017
- FDA classification
- Class II
- Brand / firm
- Ion Beam Applications S.A.
- Sold / distributed
- Worldwide Distribution - US Distribution to the states of : MA, FL, VA, PA, IL, NJ, WA, TN, LA and TX., and to the countries of: Germany, South Korea, France, Czech Republic, Italy, Poland and Sweden.
Why it was recalled
It has been observed during testing that there could be clinically non-negligible amount of dark current (a well-known physics effect observed in cyclotron) in Pencil Beam Scanning (PBS) treatment mode when system is not irradiating between layers.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Proteus 235
Get recall alerts
Free email alert whenever Ion Beam Applications S.A. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Ion Beam Applications S.A.