Ion Beam Applications S.A. recalls Proteus 235, Version PTS-11.0.1.2 and PTS-11.0.2 Product Usage: A device designed to produce and deliver a proton beam…
- Recall date
- May 16, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2424-2018
- FDA classification
- Class II
- Brand / firm
- Ion Beam Applications S.A.
- Sold / distributed
- Worldwide Distribution - US Nationwide in the states of FL and the countries of Sweden, The Netherlands
Why it was recalled
Correction vector confirmation message is lost if access point is changed after sending the correction vector. As a result, the patient will be treated in the setup position or treatment position instead of the corrected position.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Proteus 235, Version PTS-11.0.1.2 and PTS-11.0.2 Product Usage: A device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.