Medical device recalls Moderate risk

Ion Beam Applications S.A. recalls Proteus 235, Version PTS-11.0.1.2 and PTS-11.0.2 Product Usage: A device designed to produce and deliver a proton beam…

Recall date
May 16, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2424-2018
FDA classification
Class II
Brand / firm
Ion Beam Applications S.A.
Sold / distributed
Worldwide Distribution - US Nationwide in the states of FL and the countries of Sweden, The Netherlands

Why it was recalled

Correction vector confirmation message is lost if access point is changed after sending the correction vector. As a result, the patient will be treated in the setup position or treatment position instead of the corrected position.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Proteus 235, Version PTS-11.0.1.2 and PTS-11.0.2 Product Usage: A device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.