Iradimed Corporation MRidium MRI 1000 Series Infusion Sets recalled over mold contamination
- Recall date
- February 24, 2023
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Iradimed Corporation recalls Iradimed Corporation MRidium MRI 1000 Series Infusion Sets, MRI Infusuui Set - Syringe Adapter Set, REF 1057-50,
- Recall number
- Z-1310-2023
- FDA classification
- Class II
- Brand / firm
- Iradimed Corporation
- Sold / distributed
- Worldwide distribution - US Nationwide and the countries of Austria, Canada, Chile, Denmark, Germany, Great Britain, Hong Kong, Indonesia, Italy, Korea, Kuwait, Malaysia, Netherlands, Nicaragua, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Singapore, Singapore, South…
Why it was recalled
Certain Iradimed 1057 Syringe Adapter Sets, used with the MRidium MRI infusion pump, have a problem relating to a defective injection molded part which can limit the effectiveness of the syringe venting function which then can result in occlusion.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Iradimed Corporation MRidium MRI 1000 Series Infusion Sets, MRI Infusuui Set - Syringe Adapter Set, REF 1057-50,
Get recall alerts
Free email alert whenever Iradimed Corporation has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Iradimed Corporation