Ithera Medical Gmbh recalls The MSOT Acuity Echo is a Class 4 medical laser device. The MSOT Acuity is a mobile assembly of electrically-powered co…
- Recall date
- January 8, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1100-2025
- FDA classification
- Class II
- Brand / firm
- Ithera Medical Gmbh
- Sold / distributed
- Worldwide distribution: US (Nationwide) and OUS (foreign) to countries of: Austria, Germany, Italy, Netherlands, Portugal, Switzerland, and United Kingdom.
Why it was recalled
The MSOT Acuity Echo does not include an adequate instruction for safe operation of the device. The previous version of the user manuals contained a wrong value for the Nominal Ocular Hazard Distance (NOHD).
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
The MSOT Acuity Echo is a Class 4 medical laser device. The MSOT Acuity is a mobile assembly of electrically-powered components intended for imaging and analysis of soft-tissue and vasculature using combined optoacoustic (photoacoustic) and ultrasound (US) imaging.
Get recall alerts
Free email alert whenever Ithera Medical Gmbh has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Ithera Medical Gmbh