K2M product recalled over sterility concerns
- Recall date
- April 13, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- K2M, Inc recalls K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90…
- Recall number
- Z-2009-2020
- FDA classification
- Class II
- Brand / firm
- K2M, Inc
- Sold / distributed
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
Why it was recalled
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90109, UDI # 10888857108226, Size 24x6mm, 6 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.